Télécharger Imprimer la page

Rohde & Schwarz FPL1000 Manuel D'utilisation page 990

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
It is only available for PSA89600 emulation.
Parameters:
<State>
SYSTem:REVision[:STRing] <Name>
Sets the response to the REV? query to the defined string (HP emulation only, see
SYSTem:LANGuage
Parameters:
<Name>
Example:
Manual operation:
SYSTem:RSWeep <State>
Controls a repeated sweep of the E1 and MKPK HI HP model commands (for details
on the commands refer to
HP
models",
sweeping before.
This command is only available if a HP language is selected using
on page 983
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
ON | OFF | HIGH
OFF
The option is indicated as "B7J"
ON
The 40 MHz wideband is used.
The option is indicated as "B7J, 140".
HIGH
The 80 MHz wideband is used.
The option is indicated as "B7J, 122".
*RST:
OFF
on page 983).
Define the system language:
SYST:LANG '8563E'
Query the revision:
REV?
Response:
920528
Set the response to 'NewRevision':
SYST:REV:STR 'NewRevision'
Query the response:
SYST:REV:STR?
Response:
NewRevision
See
"Revision String"
Chapter 10.11.2, "Reference: GPIB commands of emulated
on page 985). If the repeated sweep is OFF, the marker is set without
ON | OFF | 1 | 0
*RST:
0
Remote commands
Emulating other instruments' commands
on page 578
SYSTem:LANGuage
990
990

Publicité

loading