Télécharger Imprimer la page

Rohde & Schwarz FPL1000 Manuel D'utilisation page 912

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
Manual operation:
CALCulate<n>:LIMit<li>:LOWer:OFFSet <Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
<li>
Parameters:
<Offset>
Manual operation:
CALCulate<n>:LIMit<li>:LOWer:SHIFt <Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
<li>
Parameters:
<Distance>
Manual operation:
CALCulate<n>:LIMit<li>:LOWer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of a lower
limit line from one horizontal point to the next.
Suffix:
<n>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
See
"X-Axis"
on page 441
Window
Limit line
Numeric value.
*RST:
0
Default unit: dB
See
"Y-Offset"
on page 439
Window
Limit line
Defines the distance that the limit line moves.
CALCulate<n>:LIMit<li>:UNIT
The unit depends on
on page 908.
Default unit: DB
See
"Shift y"
on page 442
Window
Remote commands
912
912

Publicité