R&S
FPL1000
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
Return values:
<NoiseLevel>
Example:
Usage:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe] <State>
This command turns the noise measurement at the marker position on and off.
Suffix:
<n>
<m>
Parameters:
<State>
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
INITiate<n>:CONTinuous
Window
Marker
Current noise level. The unit is the one currently active.
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK2:FUNC:NOIS ON
Switches on noise measurement for marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:FUNC:NOIS:RES?
Outputs the noise result of marker 2.
Query only
See
"Noise Measurement State"
Window
Marker
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
See
"Noise Measurement State"
See
"Switching All Noise Measurement Off"
Setting basic spectrum RF measurement parameters
on page 620.
Remote commands
on page 376
on page 376
on page 376
880
880