Parameters:
<State>
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:POINts:MINimum[:VALue] <SweepPoint>
Defines the minimum number of sweep points for the range.
Suffix:
<sb>
<ri>
Parameters:
<SweepPoint>
Example:
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:MLCalc <Function>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details see
ual ranges"
Suffix:
<sb>
Manuel d'utilisation 1179.5860.09 - 12
1..n
Power class for which the limit is defined.
ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
AND
Combines the absolute and relative limit. The limit check fails
when both limits are violated.
OR
Combines the absolute and relative limit. The limit check fails
when one of the limits is violated.
*RST:
See
"Limit Check <n>"
1 to 3
Selects the measurement range.
Minimum number of sweep points per range
Range:
*RST:
SENSe1:ESPectrum:RANGe3:POINts:MIN:VALue 400
See
" Min Sweep Points "
on page 177.
1 to 3
Sub block in a Multi-SEM measurement
Configuring and performing measurements
RELative
on page 183
1 to 32001
1
on page 184
"Limit calculation for individ-
Remote commands
672
672