Manual operation:
CALCulate<n>:LIMit<li>:STATe <State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see
on page 913).
Suffix:
<n>
<li>
Parameters:
<State>
Example:
Manual operation:
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk <State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Suffix:
<n>
<li>
<t>
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
Window
Limit line
See
"Delete Line"
on page 439
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
irrelevant
Limit line
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
See
"Disable All Lines"
Window
Limit line
Trace
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on