R&S
FPL1000
10.8.8.13
Signal count marker
The following commands control the frequency counter.
CALCulate<n>:MARKer<m>:COUNt
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
CALCulate<n>:MARKer<m>:COUNt:RESolution
CALCulate<n>:MARKer<m>:COUNt <State>
This command turns the frequency counter at the marker position on and off.
The frequency counter works for one marker only. If you perform a frequency count
with another marker, the R&S FPL1000 deactivates the frequency count of the first
marker.
The frequency counter is not available if the tracking generator is active (see
generator settings"
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
Parameters:
<State>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the frequency at the marker position.
Manuel d'utilisation 1179.5860.09 - 12
on page 300).
INITiate<n>:CONTinuous
Window
Marker
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
See
"Signal Count Marker State"
Setting basic spectrum RF measurement parameters
....................................................................................... 885
................................................................. 885
..................................................................... 886
on page 620.