INITiate<n>:CONTinuous <State>
This command controls the sweep mode for an individual channel setup.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see
Suffix:
<n>
Parameters:
<State>
<State>
Example:
Manual operation:
INITiate<n>[:IMMediate]
This command starts a (single) new measurement.
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see
Suffix:
<n>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
Remote control via
irrelevant
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
0
INIT:CONT OFF
Switches the sweep mode to single sweep.
INIT:CONT ON
Switches the sweep mode to continuous sweep.
See
"Frequency Sweep"
See
"Zero Span"
on page 103
See
"Continuous Sweep / Run Cont"
Remote control via
irrelevant
Remote commands
SCPI.
on page 103
on page 336
SCPI.