This command has the same effect as a combination of the
MAN
EXCLusive]
but via the expected power. As opposed to CALC:LIM:ESP:MODE AUTO, the power
class is not re-assigned to the input signal power dynamically, but only once when the
command is executed.
Suffix:
<n>
<li>
<sb>
Parameters:
<Power>
Example:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:COUNt <NoPowerClasses>
This command sets the number of power classes to be defined.
This command must be executed before any new power class values can be defined
using
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum.
Suffix:
<n>
<li>
<sb>
<pc>
Parameters:
<NoPowerClasses>
Example:
Manual operation:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:EXCLusive] <State>
This command selects the power class used by the measurement if
LIMit<li>:ESPectrum<sb>:MODE
Note that:
Manuel d'utilisation 1179.5860.09 - 12
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:
and the
commands; however, the power class to be used is not defined directly,
irrelevant
irrelevant
1 to 3
integer
Range:
*RST:
CALC:LIM:ESP:VAL 33
Activates manual selection of the limit line and selects the limit
line for P = 33.
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum
irrelevant
irrelevant
1 to 3
irrelevant
1 to 4
*RST:
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
See
"Adding or Removing a Power Class"
Configuring and performing measurements
-200 to 199
0
1
is set to manual.
Remote commands
CALC:LIM:ESP:MODE
and
on page 189
CALCulate<n>: