Parameters:
<DelayTime>
Example:
Manual operation:
[SENSe:]SWEep:EGATe:LENGth <GateLength>
This command defines the gate length.
Parameters:
<GateLength>
Example:
Manual operation:
[SENSe:]SWEep:EGATe:POLarity <Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.
Parameters:
<Polarity>
Example:
Manual operation:
[SENSe:]SWEep:EGATe:SOURce <Source>
This command selects the signal source for gated measurements.
If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth (10 MHz).
For more information see
Parameters:
<Source>
Example:
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
Range:
0 s to 30 s
*RST:
0 s
Default unit: S
SWE:EGAT:HOLD 100us
See
"Gate Delay"
on page 350
Range:
125 ns to 30 s
400μs
*RST:
Default unit: S
SWE:EGAT:LENG 10ms
See
"Gate Length"
on page 350
POSitive | NEGative
*RST:
POSitive
SWE:EGAT:POL POS
See
"Slope"
on page 347
"Trigger Source"
on page 344.
EXTernal | IFPower | TIME
*RST:
IFPower
SWE:EGAT:SOUR IFP
Switches the gate source to IF power.