Parameters:
<Function>
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:RLEVel <RefLevel>
This command defines the reference level for a SEM range.
In case of high speed measurements, the reference level has to be identical for all
ranges.
Suffix:
<sb>
<ri>
Parameters:
<RefLevel>
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:SWEep:TIME <SweepTime>
This command defines the sweep time for a SEM range.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
<ri>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
1..n
Selects the measurement range.
NONE | MAX | SUM
NONE
(reference ranges only:) the limit of the reference range is used;
Reference ranges always use the function "NONE".
SUM
sum of the two limit lines (calculated for linear powers) is used
MAX
maximum of the two limit lines is used
*RST:
SUM (reference range: NONE)
See
"Multi-Limit Calc <n>"
1 to 3
Sub block in a Multi-SEM measurement
1..n
Selects the measurement range.
Reference level.
Refer to the data sheet for the reference level range.
*RST:
0 dBm
Default unit: dBm
See
"Ref Level"
on page 182
1 to 3
Sub block in a Multi-SEM measurement
1..n
Selects the measurement range.
Remote commands
on page 184