R&S
FPL1000
<m>
<sb>
Parameters:
<Mode>
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult? <Measurement>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
<sb>
Parameters:
<Measurement>
Manuel d'utilisation 1179.5860.09 - 12
Marker
irrelevant
WRITe | MAXHold
WRITe
The power is calculated from the current trace.
MAXHold
The power is calculated from the current trace and compared
with the previous power value using a maximum algorithm.
See
"Power Mode"
INITiate<n>:CONTinuous
irrelevant
irrelevant
Multi-SEM: 1 to 3
for all other measurements: irrelevant
ACPower | MCACpower
ACLR measurements (also known as adjacent channel power or
multicarrier adjacent channel measurements).
Returns the power for every active transmission and adjacent
channel. The order is:
• power of the transmission channels
• power of adjacent channel (lower,upper)
• power of alternate channels (lower,upper)
The unit of the return values depends on the scaling of the y-
axis:
• logarithmic scaling returns the power in the current unit
• linear scaling returns the power in W
CN
Carrier-to-noise measurements.
Returns the C/N ratio in dB.
CN0
Carrier-to-noise measurements.
Returns the C/N ratio referenced to a 1 Hz bandwidth in
dBm/Hz.
Configuring and performing measurements
on page 143
on page 620.
Remote commands