Télécharger Imprimer la page

Rohde & Schwarz FPL1000 Manuel D'utilisation page 649

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative] <LowerLimit>[,
This command defines the relative limit of the alternate channels. The reference value
for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S FPL1000 uses
the lower value for the limit check.
Suffix:
<n>
<li>
<ch>
Parameters:
<LowerLimit>
<UpperLimit>
Manual operation:
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:RESult?
This command queries the state of the limit check for the adjacent or alternate chan-
nels in an ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<li>
<ch>
Return values:
<LowerChan>
Manuel d'utilisation 1179.5860.09 - 12
<UpperLimit>]
irrelevant
irrelevant
1..n
Alternate channel number
The limit of the lower alternate channel.
Range:
*RST:
Default unit: dB
The limit of the upper alternate channel.
Range:
*RST:
Default unit: dB
See
"Limit Check"
INITiate<n>:CONTinuous
irrelevant
irrelevant
Alternate channel number
text value
The state of the limit check for the lower alternate or adjacent
channels.
PASSED
Limit check has passed.
Configuring and performing measurements
0 dB to 100 dB
0 dB
0 dB to 100 dB
0 dB
on page 146
on page 620.
Remote commands
649
649