Parameters:
<State>
[SENSe:]AVERage<n>:TYPE <Mode>
This command selects the trace averaging mode.
Suffix:
<n>
Parameters:
<Mode>
Example:
Manual operation:
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion] <Detector>
Defines the trace detector to be used for trace analysis.
For details see
on page 395.
For EMI measurements, the trace detector is used for the initial peak search only, not
for the final test. The detector for the final test is configured using
MARKer<m>:FUNCtion:FMEasurement:DETector
If the EMI (R&S FPL1-K54) measurement option is installed and the filter type "CISPR"
is selected, additional detectors are available, even if EMI measurement is not active.
For details see
Suffix:
<n>
<t>
Parameters:
<Detector>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
Trace
ON | OFF | 1 | 0
1..n
Window
LOGarithmic
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
AVER:TYPE LIN
Switches to linear average calculation.
See
"Average Mode"
"Mapping samples to sweep points with the trace detector"
"Detectors and dwell time"
Window
Trace
APEak
Autopeak
NEGative
Negative peak
on page 406
on page 724.
on page 260.
Remote commands
CALCulate<n>: