Example:
Usage:
Manual operation:
[SENSe:]CORRection:METHod <Type>
This command selects the type of measurement to be performed with the generator.
This command is only available if the Tracking Generator is active (see
OUTPut<up>[:STATe]
Parameters:
<Type>
Example:
Manual operation:
[SENSe:]CORRection:RECall
This command restores the measurement configuration used for calibration.
This command is only available if the Tracking Generator is active (see
OUTPut<up>[:STATe]
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
INIT:CONT OFF
Selects single sweep operation
CORR:METH TRAN
Selects a transmission measurement.
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Setting only
See
"Calibrate Reflection Short"
See
"Calibrate Reflection Open"
on page 796).
REFLection
Selects reflection measurements.
TRANsmission
Selects transmission measurements.
*RST:
TRANsmission
CORR:METH TRAN
Sets the type of measurement to "transmission".
See
"Calibrate Transmission"
See
"Calibrate Reflection Short"
See
"Calibrate Reflection Open"
on page 796).
CORR:REC
See
"Recall Cal. Settings"
Remote commands
on page 302
on page 302
on page 302
on page 302
on page 302