R&S
FPL1000
SOURce<si>:INTernal:FREQuency <Frequency>
Defines the frequency of the internal generator signal.
Suffix:
<si>
Parameters:
<Frequency>
Example:
Manual operation:
SOURce<si>:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Defines the output power of the internal generator.
Suffix:
<si>
Parameters:
<Amplitude>
Example:
Manual operation:
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet <Offset>
This command defines a level offset for the external generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the external generator can be taken into
account for the setting.
Suffix:
<si>
Parameters:
<Offset>
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
irrelevant
Range:
5 kHz to 3 GHz
Increment: 0.1 Hz
*RST:
1 GHz
Default unit: HZ
SOUR:INT:FREQ 2 GHz
See
"CW Frequency"
on page 299
irrelevant
Range:
-60 dBm to +10 dBm
Increment: 0.1 dB
*RST:
-20 dBm
Default unit: DBM
SOUR:POW -30dBm
See
"Level"
on page 299
irrelevant
Range:
-200 dB to +200 dB
*RST:
0dB
Default unit: DB
//Define a level offset on the external generator
SOUR:POW:OFFS -10dB
See
"Level Offset"
on page 299
Remote commands
803
803