Télécharger Imprimer la page

Rohde & Schwarz FPL1000 Manuel D'utilisation page 651

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
Parameters:
<State>
Manual operation:
10.6.4.6
Performing an ACLR measurement
The following commands are required to perform an ACLR measurement:
CALC:MARK:FUNC:POW:SEL ACP, see
FUNCtion:POWer<sb>:SELect
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
INITiate<n>[:IMMediate]
10.6.4.7
Retrieving and analyzing measurement results
The following commands retrieve and analyze measurement results for ACLR mea-
surements.
Useful commands for channel power measurements described elsewhere
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
TRACe<n>[:DATA]
CALCulate<n>:LIMit<li>:ACPower:ACHannel:RESult?
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:RESult?
on page 643
Remote commands exclusive to channel power measurements
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult:PHZ
[SENSe:]POWer:ACHannel:MODE
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult:PHZ <State>
This command selects the unit the R&S FPL1000 returns results for power measure-
ments.
You can query results with
RESult?.
Suffix:
<n>
<m>
<sb>
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
ON | OFF | 1 | 0
*RST:
0
See
"Limit Check"
on page 146
CALCulate<n>:MARKer<m>:
on page 623
on page 620
on page 829
.......................................................................................... 646
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:
Window
Marker
irrelevant
ON | OFF | 1 | 0
Remote commands
on page 624
on page 622
on page 640
.......................................... 645
651
651

Publicité

loading