R&S
FPL1000
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO <State>
This command turns automatic selection of the input attenuation for a spurious emis-
sion measurement range on and off.
Suffix:
<ri>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe <State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
Suffix:
<ri>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue] <Gain>
This command selects the preamplification level for the range.
Suffix:
<ri>
Example:
[SENSe:]LIST:RANGe<ri>:LIMit:STARt <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
1..n
Selects the measurement range.
ON | OFF | 0 | 1
*RST:
1
LIST:RANG2:INP:ATT:AUTO ON
See
"RF Attenuation Mode"
1..n
Selects the measurement range.
ON | OFF | 1 | 0
*RST:
0
LIST:RANG2:INP:GAIN:STAT ON
See
"Preamp"
on page 212
1..n
Selects the measurement range.
LIST:RANG2:INP:GAIN 15
1..n
Selects the measurement range.
Remote commands
on page 212
692
692