CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:STATe <State>
This command turns the n dB Down marker function on and off.
Suffix:
<n>
<m>
Parameters:
<State>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
This command queries the position of the n dB down markers on the x-axis when mea-
suring in the time domain.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
Return values:
<TimeX1>
<TimeX2>
Example:
Usage:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
irrelevant
irrelevant
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
CALC:MARK:FUNC:NDBD:STAT ON
Turns the n dB Down marker on.
See
"n dB down Marker State"
INITiate<n>:CONTinuous
irrelevant
irrelevant
absolute position in time of the n dB marker to the left of the ref-
erence marker in seconds
absolute position in time of the n dB marker to the right of the
reference marker in seconds
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:TIME?
Outputs the time values of the temporary markers.
Query only
See
"n dB down Value"
Setting basic spectrum RF measurement parameters
on page 381
on page 620.