R&S
FPL1000
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STARt <Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
<ri>
<li>
Parameters:
<Level>
Example:
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STOP <Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
<ri>
<li>
Parameters:
<Level>
Example:
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
See
"Insert before Range / Insert after Range"
1 to 3
1..n
Selects the measurement range.
1..n
Power class for which the limit is defined.
Absolute limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
-13
Default unit: dBm
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STARt -10
For a detailed example see
measurement",
on page 678.
See
"Abs Limit Start / Stop <n>"
1 to 3
1..n
Selects the measurement range.
1..n
Power class for which the limit is defined.
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
-13
Default unit: dBm
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STOP -15
For a detailed example see
measurement",
on page 678.
Remote commands
on page 185
Chapter 10.6.7.10, "Example: SEM
on page 183
Chapter 10.6.7.10, "Example: SEM
667