If enabled, the trace is smoothed by the value specified using
WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
For more information see
Suffix:
<n>
<w>
<t>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]AVERage<n>:COUNt <AverageCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
Parameters:
<AverageCount>
Manual operation:
[SENSe:]AVERage<n>[:STATe<t>] <State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
"Trace smoothing"
Window
subwindow
Trace
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
DISP3:TRAC2:SMO ON
Turns on trace smoothing for trace 2 in window 3
See
"Smoothing"
on page 406
irrelevant
If you set an average count of 0 or 1, the application performs
one single sweep in single sweep mode.
In continuous sweep mode, if the average count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
See
"Sweep/Average Count"
See
"Average Count"
on page 407
Window
Remote commands
DISPlay[:
on page 402.
on page 334
on page 814.
821