Example:
Manual operation:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>
<w>
<t>
Parameters:
<Offset>
Example:
Manual operation:
10.8.3.2
Configuring the attenuation
INPut<ip>:ATTenuation
INPut<ip>:EATT:AUTO
INPut<ip>:ATTenuation:AUTO
INPut<ip>:ATTenuation <Attenuation>
This command defines the total attenuation for RF input.
If you set the attenuation manually, it is no longer coupled to the reference level, but
the reference level is coupled to the attenuation. Thus, if the current reference level is
not compatible with an attenuation that has been set manually, the command also
adjusts the reference level.
Suffix:
<ip>
Parameters:
<Attenuation>
Manuel d'utilisation 1179.5860.09 - 12
DISP:TRAC:Y:RLEV -60dBm
See
"Reference Level"
<Offset>
irrelevant
subwindow
Not supported by all applications
irrelevant
Range:
*RST:
Default unit: DB
DISP:TRAC:Y:RLEV:OFFS -10dB
See
"Shifting the Display (Offset)"
See
"Shifting the Display (Offset)"
............................................................................................................ 773
............................................................................................................ 774
................................................................................................. 774
1 | 2
irrelevant
Range:
Increment: 5 dB (with optional electr. attenuator: 1 dB)
*RST:
Default unit: DB
Setting basic spectrum RF measurement parameters
on page 319
-200 dB to 200 dB
0dB
see data sheet
10 dB (AUTO is set to ON)
Remote commands