Example:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:MODE <Mode>
Which limit line is to be used for an SEM measurement depends on the power class
the input signal power belongs to. This command defines wether the power class is
determined automatically or manually.
Suffix:
<n>
<li>
<sb>
Parameters:
<Mode>
Example:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:VALue <Power>
This command activates the manual limit line selection as and specifies the expected
power as a value. Depending on the entered value, the associated predefined limit
lines is selected.
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
Defines the value range for power class 3 as <Max2> to
<Max3>.
The value range for power class 4 is defined as <Max3> to
200.
Only available for CALC:LIM:ESP:PCL:COUNT = 4
Range:
-199 to + 199, <Max3> must be higher than
<Max2>
CALC:LIM:ESP:LIM -50,50,70
Defines the following power classes:
<-200, -50>
<-50, 50>
<50, 70>
<70, 200>
Query:
CALC:LIM:ESP:LIM?
Response:
-200,-50,50,70,200
irrelevant
irrelevant
1 to 3
AUTO
The power class (and thus the limit line) is assigned dynamically
according to the currently measured channel power.
MANUAL
One of the specified power classes is selected manually for the
entire measurement. The selection is made with the
CALCulate<n>:LIMit<li>:ESPectrum<sb>:
PCLass<pc>[:EXCLusive]
*RST:
AUTO
CALC:LIM:ESP:MODE AUTO
Activates automatic selection of the limit line.
Remote commands
command.