Rohde & Schwarz FPL1000 Manuel D'utilisation page 664

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STARt <Frequency>
This command defines the start frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the SEM sweep list covers, the R&S FPL1000 will not mea-
sure the ranges that are outside the span - results may be invalid.
greater than the span the SEM sweep list covers, the R&S FPL1000 will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
For more information see
Suffix:
<sb>
<ri>
Parameters:
<Frequency>
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STOP <Frequency>
This command defines the stop frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the SEM sweep list covers, the R&S FPL1000 will not mea-
sure the ranges that are outside the span - results may be invalid.
greater than the span the SEM sweep list covers, the R&S FPL1000 will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S FPL1-K54) option
Return value for query is always PULS.
MIL
MIL Std (6 dB) - requires EMI (R&S FPL1-K54) option
P5
5 Pole filters
*RST:
NORM
Refer to the datasheet for available filter bandwidths.
See
"Filter Type"
on page 181
"Ranges and range settings"
1..n
1..n
Selects the measurement range.
Numeric value. Note that the minimum frequency range of a
SEM range is 20 Hz.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
See
"Range Start / Range Stop"
Remote commands
on page 170.
on page 181
664
664