[SENSe:]SWEep:TIME <Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths. Note that this command queries only the time
required to capture the data, not to process it. To obtain an estimation of the total cap-
ture and processing time, use the
Parameters:
<Time>
Manual operation:
[SENSe:]SWEep:TIME:AUTO <State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State>
Example:
Manual operation:
[SENSe:]SWEep:TYPE <Type>
This command selects the sweep type.
The R&S FPL1000 automatically sets the optimal sweep type for the current measure-
ment (sweep or FFT). The sweep type cannot be defined manually, except for EMI
measurements. For measurements other than EMI, the sweep type is set to "Auto"
automatically.
Parameters:
<Type>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
[SENSe:]SWEep:DURation?
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
See
" Sweep Time "
on page 144
See
" Sweep Time "
on page 332
See
"Sweep Time"
on page 344
ON | OFF | 0 | 1
*RST:
1
SWE:TIME:AUTO ON
Activates automatic sweep time.
See
"Harmonic Sweep Time"
See
" Sweep Time "
on page 332
See
"Default Coupling"
AUTO
Automatic selection of the sweep type between sweep mode
and FFT.
FFT
FFT mode