R&S
FPL1000
Example:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult<t>?
This command queries the results of the AM modulation depth measurement..
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<m>
<t>
Return values:
<ModulationDepth>
Usage:
10.6.13.2
Example: measuring the AM modulation depth
This example demonstrates how to determine the AM modulation depth in a remote
environment. Note that without a real input signal this measurement will not return use-
ful results.
//-----------Configuring the measurement ------------
*RST
//Reset the instrument
FREQ:CENT 100MHz
//Set center frequency
FREQ:SPAN 10KHz
// Set span
CALC:MARK:FUNC:MDEP ON
//Activate AM modulation depth measurement.
//--------------Performing the Measurement-----
INIT:CONT OFF
//Selects single sweep mode.
INIT:IMM
// Perform a single measurement
CALC:MARK:FUNC:MDEP:SEAR ONCE
//Initiates a search for signals in the current trace.
//---------------Retrieving Results-------------
Manuel d'utilisation 1179.5860.09 - 12
CALC:MARK:FUNC:MDEP:SEAR ONCE
Executes the search of an AM modulated signal at the currently
available trace.
See
"Search Signals"
INITiate<n>:CONTinuous
Window
Marker
Trace
Modulation depth in %.
Query only
Configuring and performing measurements
on page 255
on page 620.
Remote commands
728
728