Parameters:
<Factor>
Example:
Manual operation:
[SENSe:]FREQuency:OFFSet <Offset>
This command defines a frequency offset.
If this value is not 0 Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also
Parameters:
<Offset>
Example:
Manual operation:
[SENSe:]FREQuency:SPAN <Span>
This command defines the frequency span.
If you set a span of 0 Hz in the Spectrum application, the R&S FPL1000 starts a mea-
surement in the time domain.
Parameters:
<Span>
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
1 to 100 PCT
*RST:
Default unit: PCT
//Couple frequency step size to span and define a step size fac-
tor
FREQ:CENT:STEP:LINK SPAN
FREQ:CENT:STEP:LINK:FACT 20PCT
See
"Center Frequency Stepsize"
"Frequency Offset"
on page 313.
Range:
*RST:
Default unit: HZ
FREQ:OFFS 1GHZ
See
"Frequency Offset"
The minimum span for measurements in the frequency domain
is 10 Hz.
For SEM and spurious emission measurements, the minimum
span is 20 Hz.
Range:
*RST:
Default unit: Hz
See
"Zero Span"
See
"Span"
See
"Zero Span"
See
"Last Span"
Setting basic spectrum RF measurement parameters
10
-1 THz to 1 THz
0 Hz
on page 313
0 Hz to fmax
Full span
on page 103
on page 311
on page 312
on page 312
Remote commands
on page 313
763
763