Parameters:
<Percentage>
Example:
Manual operation:
[SENSe:]PMETer<p>:FREQuency <Frequency>
This command defines the frequency of the power sensor.
Suffix:
<p>
Parameters:
<Frequency>
Example:
Manual operation:
[SENSe:]PMETer<p>:FREQuency:LINK <Coupling>
This command selects the frequency coupling for power sensor measurements.
Suffix:
<p>
Parameters:
<Coupling>
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
Range:
0.001 to 99.999
*RST:
99.999
Default unit: %
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
See
"Duty Cycle"
on page 289
Power sensor index
The available value range is specified in the data sheet of the
power sensor in use.
*RST:
50 MHz
Default unit: HZ
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
See
"Frequency Manual"
Power sensor index
CENTer
Couples the frequency to the center frequency of the analyzer
MARKer1
Couples the frequency to the position of marker 1
OFF
Switches the frequency coupling off
*RST:
CENTer
PMET2:FREQ:LINK CENT
Couples the frequency to the center frequency of the analyzer
See
"Frequency Coupling"
Remote commands
on page 287
on page 287
798
798