R&S
FPL1000
Example:
Manual operation:
INPut<ip>:EATT:AUTO <State>
INPut<ip>:ATTenuation:AUTO <State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S FPL1000 determines the signal level for opti-
mal internal data processing and sets the required attenuation accordingly.
For the R&S FPL1000, these commands are identical.
Suffix:
<ip>
Parameters:
<State>
Example:
Manual operation:
10.8.3.3
Configuring a preamplifier
INPut<ip>:GAIN:STATe
INPut<ip>:GAIN:STATe <State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
If activated, the input signal is amplified by 20 dB. The preamplifier is only active below
3 GHz (R&S FPL1003) or 7.5 GHz (R&S FPL1007).
Suffix:
<ip>
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
See
"Attenuation Mode / Value"
1 | 2
irrelevant
ON | OFF | 0 | 1
*RST:
1
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
See
"Attenuation Mode / Value"
................................................................................................................... 774
1 | 2
irrelevant
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Remote commands
on page 321
on page 321