10.6.14.2
Configuring EMI markers
The commands required to configure EMI markers are described here.
Useful commands for configuring EMI markers described elsewhere:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
CALCulate<n>:DELTamarker<m>:MREFerence
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
CALCulate<n>:DELTamarker<m>:LINK
CALCulate<n>:MARKer<m>:TRACe
Remote commands exclusive to configuring EMI markers
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector <Detector>
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector <Detector>
This command selects the detector for a specific marker during the final measurement.
If the marker is not yet active, the command also turns the marker on.
Suffix:
<n>
<m>
Parameters:
<Detector>
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
on page 844
Window
Marker
OFF
no final measurement is performed
AVER
average detector
CAV
CISPR Average detector
CRMS
RMS Average detector
POS
maximum peak detector
QPE
quasipeak detector
*RST:
OFF
See
"Final Test Detector"
Remote commands
on page 843
on page 841
on page 843
on page 841
on page 840
on page 843
on page 838
................................. 724
.......................................... 724
on page 271
on page 839
730