R&S
FPL1000
10.6.3.1
Configuring the compression point measurement
The following commands are required to configure compression point measurements
using an internal tracking generator.
For details see
For a programming example using remote commands, see
gramming example: measuring compression
Useful commands for compression point measurements described elsewhere:
●
"Power sweep commands"
●
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:
RPOSition
Remote commands exclusive to compression point measurements:
CONFigure:CMEasurement:REFerence:VALue
CONFigure:CMEasurement:REFerence[:MODE]
CONFigure:CMEasurement:RESult:CVALue
CONFigure:CMEasurement:RESult:P1DB[:STAT]
CONFigure:CMEasurement:RESult:P3DB[:STAT]
CONFigure:CMEasurement:RESult:PNDB[:STAT]
CONFigure:CMEasurement[:STATe]
FETCh:CMEasurement:P1DB[:RESult]?
FETCh:CMEasurement:P3DB[:RESult]?
FETCh:CMEasurement:PNDB[:RESult]?
CONFigure:CMEasurement:REFerence:VALue <ReferenceValue>
Defines the reference value used to calculate the compression point for
CMEasurement:REFerence[:MODE]
Parameters:
<ReferenceValue>
Example:
Manual operation:
CONFigure:CMEasurement:REFerence[:MODE] <Mode>
Determines how the reference value for the compression point measurement is
defined.
Parameters:
<Mode>
Manuel d'utilisation 1179.5860.09 - 12
Chapter 6.2.3, "Power sweep
on page 801
on page 776
....................................................................................... 628
.................................................................................. 629
.................................................................................. 629
................................................................................. 629
Default unit: DBM
CONF:CME:REF MAN
CONF:CME:REF:VAL -30
See
"Reference"
AUTO | MANual
AUTO
The reference value is determined automatically.
Configuring and performing measurements
measurements",
Chapter 10.6.3.2, "Pro-
points",
on page 629.
...................................................................... 626
.................................................................... 626
........................................................................... 627
................................................................... 627
................................................................... 628
.................................................................. 628
MANual.
on page 131
Remote commands
on page 125.
CONFigure:
632
632