Parameters:
<Frequency>
<RBW>
<MeasTime>
<TriggerSource>
<TriggerLevel>
<TriggerOffset>
<Detector>
<OfPulses>
Return values:
<PowerLevel>
[SENSe:]MPOWer:RESult:MIN?
This command queries the lowest pulse power that has been measured during a pulse
power measurement.
If there are no results, the command returns an error.
Return values:
<PulsePower>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
Defines the pulse frequency.
Range:
0 to Fmax
Default unit: Hz
Defines the resolution bandwidth.
Default unit: HZ
Defines the measurement time.
Range:
1 µs to 30 s
Default unit: S
Selects a trigger source.
Defines a trigger level.
The trigger level is available for the video trigger. In that case,
the level is a percentage of the diagram height.
In case of an external trigger, the R&S FPL1000 uses a fix TTL
level.
Range:
0 to 100
Default unit: PCT
Defines the trigger delay.
Range:
0 s to 30 s
*RST:
0 s
Default unit: s
Selects the detector and therefore the way the measurement is
evaluated.
MEAN
Calculates the RMS pulse power.
PEAK
Calculates the peak pulse power.
Defines the number of pulses included in the measurement.
Range:
1 to 32001
Pulse power level.
The result is a list of floating point values separated by commas.
The unit is dBm.
Lowest power level of the pulse power measurement.
The unit is dBm.