R&S
FPL1000
Example:
10.8.2.2
Configuring the sweep
Useful commands for configuring sweeps described elsewhere:
●
[SENSe:]AVERage<n>:COUNt
●
[SENSe:]AVERage<n>[:STATe<t>]
●
[SENSe:]AVERage<n>:TYPE
Remote commands exclusive to configuring sweeps:
[SENSe:]SWEep:COUNt
[SENSe:]SWEep:DURation?
[SENSe:]SWEep:FFTSubspan?
[SENSe:]SWEep:OPTimize
[SENSe:]SWEep[:WINDow<n>]:POINts
[SENSe:]SWEep:TIME
[SENSe:]SWEep:TIME:AUTO
[SENSe:]SWEep:TYPE
[SENSe:]SWEep:TYPE:USED
[SENSe:]SWEep:COUNt <SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In continuous sweep mode, the application calculates the moving average over the
average count.
In single sweep mode, the application stops the measurement and calculates the aver-
age after the average count has been reached.
Parameters:
<SweepCount>
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
*RST:
LOGarithmic
BAND:VID:TYPE LIN
Logarithmic amplifier in front of the video filter is disabled.
on page 815
on page 816
.......................................................................................................... 767
.................................................................................................... 768
............................................................................................... 768
...................................................................................................... 768
................................................................................... 769
............................................................................................................. 770
.................................................................................................. 770
............................................................................................................ 770
................................................................................................. 771
When you set a sweep count of 0 or 1, the R&S FPL1000 per-
forms one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
See
"Sweep/Average Count"
Remote commands
on page 815
on page 334
773
773