R&S
FPL1000
Suffix:
<t>
Parameters:
<Length>
Example:
Manual operation:
[SENSe:]SWEep:EGATe:TRACe<t>:STARt<gr> <Time>
This command defines the start time for a gate range.
Suffix:
<t>
<gr>
Parameters:
<Time>
Example:
Manual operation:
[SENSe:]SWEep:EGATe:TRACe<t>[:STATe<gr>] <State>
This command includes or excludes a gate range for a particular trace.
Suffix:
<t>
<gr>
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
irrelevant
Range:
100 ns to 1000 s
*RST:
2 ms
Default unit: s
SWE:EGAT:TRAC:PER 5ms
Defines the period for gated triggering to 5 ms.
See
"Period"
on page 225
Trace
1..n
gate range
The value range depends on the gate period you have set for
the selected trace with
PERiod. The following rules apply:
• the start time may not be higher than the length of the gate
• the start time may not be lower than the stop time of the gate
range of a lower order
The reset values depend on the gate range.
• for gate range 1, the start time is 0 ms
• for gate range 3, the start time is 2 ms
• for gate range 5, the start time is 4 ms
Default unit: s
SWE:EGAT:TRAC1:STAR1 3ms
Sets the Starting point for range 1 on trace 1 at 3 ms.
See
"Range <x> Start/Stop"
Trace
gate range
ON | OFF | 0 | 1
OFF | 0
Switches the function off
Remote commands
[SENSe:]SWEep:EGATe:TRACe<t>:
on page 226
703
703