CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage <State>
This command switches on or off averaging for the active power measurement in zero
span in the window specified by the suffix <n>. If activated, a time domain value is cal-
culated from the trace after each sweep; in the end, all values are averaged to calcu-
late the final result.
The number of results required for the calculation of average is defined with
]AVERage<n>:COUNt
Averaging is reset by switching it off and on again.
Synchronization to the end of averaging is only possible in single sweep mode.
Suffix:
<n>
<m>
Parameters:
<State>
Example:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd <State>
This command switches on or off the peak-hold function for the active power measure-
ment in zero span in the window specified by the suffix <n>. If activated, the peak for
each sweep is compared to the previously stored peak; the maximum of the two is
stored as the current peak.
The peak-hold function is reset by switching it off and on again.
Suffix:
<n>
<m>
Parameters:
<State>
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary[:STATe] <State>
This command turns time domain power measurements on and off. This measurement
is only available in zero span.
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
.
Window
Marker
ON | OFF | 1 | 0
*RST:
0
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:AVER ON
Switches on the calculation of average.
AVER:COUN 200
Sets the measurement counter to 200.
INIT;*WAI
Starts a sweep and waits for the end.
Window
Marker
ON | OFF | 1 | 0
*RST:
0
Remote commands
[SENSe:
711