R&S
FPL1000
For more information see
Suffix:
<sb>
<ri>
Parameters:
<Frequency>
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation <Attenuation>
This command defines the input attenuation for a SEM range.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
<ri>
Parameters:
<Attenuation>
Manual operation:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation:AUTO <State>
This command turns automatic selection of the input attenuation for a SEM range on
and off.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
<ri>
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
"Ranges and range settings"
1..n
1..n
Selects the measurement range.
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
See
"Range Start / Range Stop"
1 to 3
1..n
Selects the measurement range.
Numeric value.
Refer to the data sheet for the attenuation range.
*RST:
10 dB
Default unit: dB
See
"RF Attenuation"
on page 182
1 to 3
1..n
Selects the measurement range.
ON | OFF | 0 | 1
*RST:
1
Remote commands
on page 170.
on page 181
665
665