[SENSe:]POWer:ACHannel:FILTer[:STATe][:ALL]
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel <Alpha>
This command defines the roll-off factor for the adjacent channel weighting filter.
Parameters:
<Alpha>
Manual operation:
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL] <Value>
This command defines the alpha value for the weighting filter for all channels.
Parameters:
<Value>
Example:
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch> <Alpha>
This command defines the roll-off factor for the alternate channel weighting filter.
Suffix:
<ch>
Parameters:
<Alpha>
Manual operation:
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch> <Alpha>
This command defines the roll-off factor for the transmission channel weighting filter.
Suffix:
<ch>
Parameters:
<Alpha>
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
......................................................................... 637
Roll-off factor
Range:
0 to 1
*RST:
0.22
See
"Weighting Filters"
*RST:
0.22
POW:ACH:FILT:ALPH:ALL 0.35
1..n
Alternate channel number
Roll-off factor
Range:
0 to 1
*RST:
0.22
See
"Weighting Filters"
1..n
Tx channel number
Roll-off factor
Range:
0 to 1
*RST:
0.22
See
"Weighting Filters"