Switching to manual mode is only possible via remote control then. This function is
only available for the GPIB interface.
5 .
To enable the keys of the R&S FPL1000 again, switch the instrument to local mode
(GTL - Go to Local), i.e. deactivate the REN line of the remote control interface.
If the instrument is operated exclusively in remote control, it is recommended that you
switch off the display. For details see
9.5.9
How to return to manual operation
Before you switch back to manual operation, all remote command processing must be
completed. Otherwise, the instrument will switch back to remote control immediately.
► Select the "Local" softkey, or use the following GPIB command:
status = viGpibControlREN(vi, VI_GPIB_REN_ADDRESS_GTL)
If you select the "Local" softkey while a self-alignment or a self-test is still running
(which was started remotely), the instrument only returns to the manual operation state
when the alignment or test is completed.
Manuel d'utilisation 1179.5860.09 - 12
Network operation and remote control
How to set up a network and remote control
"Remote Display Update"
on page 575.