Télécharger Imprimer la page

Rohde & Schwarz FPL1000 Manuel D'utilisation page 567

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
Bit No.
2
3
4
5
6
7
9.2.2.4
STATus:OPERation register
The STATus:OPERation register contains information on current activities of the
R&S FPL1000. It also contains information on activities that have been executed since
the last read out.
You can read out the register with
OPERation[:EVENt]?.
Table 9-4: Meaning of the bits used in the STATus:OPERation register
Bit No.
0
1-2
3
4
5
Manuel d'utilisation 1179.5860.09 - 12
Manuel d'utilisation 1179.5860.09 – 12
Meaning
Query Error
This bit is set if either the controller wants to read data from the instrument without having sent a
query, or if it does not fetch requested data and sends new instructions to the instrument
instead. The cause is often a query which is faulty and hence cannot be executed.
Device-dependent Error
This bit is set if a device-dependent error occurs. An error message with a number between
-300 and -399 or a positive error number, which denotes the error in greater detail, is entered
into the error queue.
Execution Error
This bit is set if a received command is syntactically correct but cannot be performed for other
reasons. An error message with a number between -200 and -300, which denotes the error in
greater detail, is entered into the error queue.
Command Error
This bit is set if a command is received, which is undefined or syntactically incorrect. An error
message with a number between -100 and -200, which denotes the error in greater detail, is
entered into the error queue.
User Request
This bit is set when the instrument is switched over to manual control.
Power On (supply voltage on)
This bit is set on switching on the instrument.
Meaning
CALibrating
This bit is set as long as the instrument is performing a self-alignment.
Not used
SWEeping
Sweep is being performed in base unit (applications are not considered); identical to bit 4
In applications, this bit is not used.
MEASuring
Measurement is being performed in base unit (applications are not considered); identical to bit 3
In applications, this bit is not used.
Waiting for TRIgger
Instrument is ready to trigger and waiting for trigger signal.
Network operation and remote control
STATus:OPERation:CONDition?
Status reporting system
STATus:
or
567
567

Publicité

loading