R&S
FPL1000
Parameters:
<Bandwidth>
Example:
Manual operation:
[SENSe:]BWIDth:VIDeo:AUTO <State>
[SENSe:]BANDwidth:VIDeo:AUTO <State>
This command couples and decouples the video bandwidth to the resolution band-
width.
Parameters:
<State>
Example:
Manual operation:
[SENSe:]BWIDth:VIDeo:RATio <Ratio>
[SENSe:]BANDwidth:VIDeo:RATio <Ratio>
This command defines the coupling ratio of the video bandwidth to the resolution band-
width (VBW/RBW).
Parameters:
<Ratio>
Example:
Manual operation:
[SENSe:]BWIDth:VIDeo:TYPE <Mode>
[SENSe:]BANDwidth:VIDeo:TYPE <Mode>
This command enables or disables the logarithmic amplifier in front of the video filter in
the signal path.
Parameters:
<Mode>
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
refer to data sheet
*RST:
AUTO is set to ON
Default unit: HZ
BAND:VID 10 kHz
See
"VBW"
on page 332
ON | OFF | 0 | 1
*RST:
1
BAND:VID:AUTO OFF
See
"VBW"
on page 332
See
"RBW/VBW"
on page 333
See
"Default Coupling"
Range:
0,001 to 1000
*RST:
1
BAND:VID:RAT 3
Sets the video bandwidth to 3*resolution bandwidth.
See
"RBW/VBW"
on page 333
LINear
The logarithmic amplifier in front of the video filter is bypassed to
process linear detector samples.
LOGarithmic
The logarithmic amplifier in front of the video filter is enabled to
process logarithmic detector samples.