FPL1000
SOURce<si>:POWer:STOP <StopLevel>
Defines the end power level of the internal generator signal for power sweeps. This
value must be higher than the start level (see
on page 802).
Suffix:
<si>
Parameters:
<StopLevel>
Example:
Manual operation:
Programming example for a tracking generator
The following example demonstrates how to work with a tracking generator in a remote
environment. It requires the Tracking Generator option (R&S FPL1-B9) to be installed
on the R&S FPL1000.
//--------------Preparing the instrument -----------
//Reset the instrument
*RST
//Set the frequency span.
SENS:FREQ:STAR 10HZ
SENS:FREQ:STOP 1MHZ
//--------------Configuring the use of the internal generator -----------
OUTP ON
SOUR:FREQ:COUP ON
//Set the generator output level to -10 dBm.
SOUR:POW -10DBM
//Define a frequency offset of 100kHz
SOUR:INT:FREQ:OFFS 100KHZ
//--------------Performing the calibration measurement -----------
//Perform a transmission measurement with direct connection between the generator
//and the analyzer and wait till the end
SENS:CORR:METH TRAN
SENS:CORR:COLL:ACQ THR; *WAI
//--------------Retrieving the calibration trace results -----------
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
irrelevant
Range:
-60 to +10
*RST:
-20
Default unit: DBM
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
See
"Level Start / Level Stop"
Remote commands
SOURce<si>:POWer:STARt
on page 305