[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer <Threshold>
When the reference level is adjusted automatically using the
LEVel
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines a lower threshold the signal must fall
below (compared to the last measurement) before the reference level is adapted auto-
matically.
Parameters:
<Threshold>
Example:
Manual operation:
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer <Threshold>
When the reference level is adjusted automatically using the
LEVel
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines an upper threshold the signal must
exceed (compared to the last measurement) before the reference level is adapted
automatically.
Parameters:
<Threshold>
Example:
Example:
Manual operation:
[SENSe:]ADJust:CONFigure:TRIGger <State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON, for example).
See
Parameters:
<State>
Manuel d'utilisation 1179.5860.09 - 12
on page 761 command, the internal attenuators and the preamplifier are also
Range:
*RST:
Default unit: dB
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18 dBm.
See
on page 761 command, the internal attenuators and the preamplifier are also
Range:
*RST:
Default unit: dB
SENS:ADJ:CONF:HYST:UPP 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22 dBm.
See
"Adjusting settings automatically during triggered measurements"
ON | OFF | 0 | 1
OFF | 0
Switches the function off
Setting basic spectrum RF measurement parameters
0 dB to 200 dB
+1 dB
"Lower Level Hysteresis"
0 dB to 200 dB
+1 dB
"Upper Level Hysteresis"
Remote commands
[SENSe:]ADJust:
on page 354
[SENSe:]ADJust:
on page 353
on page 352.