R&S
FPL1000
Example:
Usage:
LAYout:WINDow<n>:ADD? <Direction>,<WindowType>
This command adds a measurement window to the display. Note that with this com-
mand, the suffix <n> determines the existing window next to which the new window is
added, as opposed to LAYout:ADD[:WINDow]?, for which the existing window is
defined by a parameter.
To replace an existing window, use the
This command is always used as a query so that you immediately obtain the name of
the new window as a result.
Suffix:
<n>
Query parameters:
<Direction>
<WindowType>
Return values:
<NewWindowName> When adding a new window, the command returns its name (by
Example:
Usage:
LAYout:WINDow<n>:IDENtify?
This command queries the name of a particular display window (indicated by the <n>
suffix) in the active channel setup.
Note: to query the index of a particular window, use the
WINDow]?
Manuel d'utilisation 1179.5860.09 - 12
LAY:SPL 1,4,70
Moves the splitter between window 1 ('Frequency Sweep') and 3
('"Marker Peak List"') towards the top (70%) of the screen.
The following commands have the exact same effect, as any
combination of windows above and below the splitter moves the
splitter vertically.
LAY:SPL 3,2,70
LAY:SPL 4,1,70
LAY:SPL 2,1,70
Setting only
Window
LEFT | RIGHt | ABOVe | BELow
Type of measurement window you want to add.
LAYout:ADD[:WINDow]?
See
ble window types.
default the same as its number) as a result.
LAY:WIND1:ADD? LEFT,MTAB
Result:
'2'
Adds a new window named '2' with a marker table to the left of
window 1.
Query only
command.
Remote commands for result displays
LAYout:WINDow<n>:REPLace
on page 743 for a list of availa-
LAYout:IDENtify[:
Remote commands
command.
754
754