If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
<m>
Example:
Manual operation:
CALCulate<n>:UNIT:POWer <Unit>
UNIT<n>:POWer <Unit>
This command selects the unit of the y-axis.
The unit applies to all power-based measurement windows with absolute values.
Suffix:
<n>
Parameters:
<Unit>
Example:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel
This command defines the reference level (for all traces in all windows).
With a reference level offset ≠ 0, the value range of the reference level is modified by
the offset.
Suffix:
<n>
<w>
<t>
Parameters:
<ReferenceLevel>
Manuel d'utilisation 1179.5860.09 - 12
Window
Marker
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
See
"Reference Level = Marker Level"
irrelevant
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
(Units based on 1 MHz require installed R&S FPL1-K54 (EMI
measurements) option.)
*RST:
UNIT:POW DBM
Sets the power unit to dBm.
<ReferenceLevel>
irrelevant
subwindow
Not supported by all applications
irrelevant
The unit is variable.
Range:
*RST:
Default unit: DBM
Setting basic spectrum RF measurement parameters
dBm
see datasheet
0 dBm
Remote commands
on page 371