R&S
FPL1000
Bit No.
6-7
8
9
10-14
15
9.2.2.5
STATus:QUEStionable register
The STATus:QUEStionable register contains information on instrument states that do
not meet the specifications.
The STAT:QUES:SYNC register is used by the applications and is thus described in the
individual applications' User Manuals.
You can read out the register with STAT:QUES:COND or STAT:QUES:EVEN.
The STATus:QUEStionable register "sums up" the information from all subregisters
(e.g. bit 2 sums up the information for all STATus:QUEStionable:TIMe registers).
For some subregisters, there may be separate registers for each active channel. Thus,
if a status bit in the STATus:QUEStionable register indicates an error, the error may
have occurred in any of the channel-specific subregisters. In this case, you must check
the subregister of each channel to determine which channel caused the error. By
default, querying the status of a subregister always returns the result for the currently
selected channel.
Table 9-5: Meaning of the bits used in the STATus:QUEStionable register
Bit No.
0
1
2
3
Manuel d'utilisation 1179.5860.09 - 12
Meaning
Not used
HardCOPy in progress
This bit is set while the instrument is printing a hardcopy.
not used
Not used
This bit is always 0.
Meaning
"EXTended"
This bit indicates further status information not covered by the other status registers in any of
the active channels.
"Battery LOW"
This bit is set if the battery needs to be charged or a fixed power supply must be connected to
the R&S FPL1000 to continue operation.
"TIMe"
This bit is set if a time error occurs in any of the active channels.
The
STATus:QUEStionable:TIMe register
"POWer"
This bit is set if the measured power level in any of the active channels is questionable.
The
STATus:QUEStionable:POWer register
Network operation and remote control
Status reporting system
provides more information on the error type.
provides more information on the error type.