R&S
FPL1000
Parameters:
<Level>
Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>:LIMit:STATe <State>
This command turns the limit check for all spurious emission measurement ranges on
and off.
Suffix:
<ri>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>:LIMit:STOP <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
Parameters:
<Level>
Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>:POINts[:VALue] <Points>
This command defines the number of sweep points in a spurious emission measure-
ment range.
Suffix:
<ri>
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
Absolute limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
LIST:RANG2:LIM:STAR 200
See
"Abs Limit Start/ Abs Limit Stop"
irrelevant
ON | OFF | 1 | 0
*RST:
0
LIST:RANG2:LIM:STAT ON
See
"Limit Check"
on page 213
1..n
Selects the measurement range.
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
LIST:RANG2:LIM:STOP 200
See
"Abs Limit Start/ Abs Limit Stop"
1..n
Selects the measurement range.
Remote commands
on page 213
on page 213
693
693