Rohde & Schwarz FPL1000 Manuel D'utilisation page 698

Masquer les pouces Voir aussi pour FPL1000:

Publicité

Les langues disponibles

Les langues disponibles

®
R&S
FPL1000
10.6.8.4
Adjusting the X-axis to the range definitions
[SENSe:]LIST:XADJust
Sets the x-axis range for the spurious emission measurement from the start frequency
of the first sweep range to the stop frequency of the last sweep range.
Example:
Usage:
10.6.8.5
Performing a spurious measurement
The following commands are required to perform a Spurious measurement:
SENS:SWE:MODE LIST, see
INITiate<n>[:IMMediate]
surements",
10.6.8.6
Retrieving and saving settings and results
The following commands analyze and retrieve measurement results for Spurious mea-
surements.
Useful commands for spurious emission measurements described elsewhere
CALCulate<n>:LIMit<li>:FAIL?
TRACe<n>[:DATA]
TRACe<n>[:DATA]:MEMory?
TRACe<n>[:DATA]:X?
10.6.8.7
Programming example: spurious emissions measurement
In the following example, the Spurious Emissions measurement is configured by defin-
ing ranges and parameters to create the following sweep list.
Note that this example is primarily meant to demonstrate the remote control com-
mands, it does not necessarily reflect a useful measurement task.
//------------Preparing the measurement-------------
*RST
//Resets the instrument
SWE:MODE LIST
//Activates spurious emissions measurement
INIT:CONT OFF
//Selects single sweep mode.
//Spurious measurement has to be in single sweep mode to be configured
Manuel d'utilisation 1179.5860.09 - 12
SENS:LIST:XADJ
Event
[SENSe:]SWEep:MODE
on page 620, see
on page 618
on page 829
on page 831
Configuring and performing measurements
on page 653
Chapter 10.6.1, "Performing mea-
on page 915
on page 830
Remote commands
698
698