CALCulate<n>:LIMit<li>:UPPer:STATe <State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with
LIMit<li>:NAME
Suffix:
<n>
<li>
Parameters:
<State>
Manual operation:
CALCulate<n>:LIMit<li>:UPPer:THReshold <Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S FPL1000 uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
<li>
Parameters:
<Limit>
Manual operation:
10.8.10.2
Managing limit lines
Useful commands for managing limit lines described in the R&S FPL1000 User Man-
ual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
Manuel d'utilisation 1179.5860.09 - 12
Setting basic spectrum RF measurement parameters
on page 908.
irrelevant
Limit line
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
See
"Visibility"
on page 438
irrelevant
Limit line
Numeric value.
The unit depends on
on page 908.
*RST:
-200
Default unit: dBm
See
"Threshold"
on page 440
Remote commands
CALCulate<n>:LIMit<li>:UNIT
CALCulate<n>:
917