R&S
FPL1000
If the number of error messages exceeds the capacity of the error buffer, the oldest
error message is removed before the newest one is inserted. To clear the message
buffer use the "Clear Error List" button. It is automatically cleared when the
R&S FPL1000 is shut down.
The following information is available:
No
Error
Date/Time
Remote command:
SYSTem:ERRor:LIST?
Clear Error List
Deletes the error message buffer for remote operation.
Note: The remote error list is automatically cleared when the R&S FPL1000 is shut
down.
Remote command:
SYSTem:ERRor:CLEar:REMote
9.4.6
Returning to manual mode ("Local")
When switched on, the instrument is always in the manual measurement mode and
can be operated via the front panel. As soon as the instrument receives a remote com-
mand, it is switched to the remote control mode.
In remote control mode, all keys of the instrument except the [PRESET] key are dis-
abled. The "LOCAL" softkey and the
Local
The instrument switches from remote to manual operation.
Note:
●
If the local lockout function (LLO) has been activated in the remote control mode,
manual operation is no longer available until GTL is executed.
●
Before you switch back to manual operation, all remote command processing must
be completed. Otherwise, the instrument will switch back to remote control immedi-
ately.
●
If you select the "Local" softkey while a self-alignment or a self-test is still running
(which was started remotely), the instrument only returns to the manual operation
state when the alignment or test is completed.
Furthermore, when you return to manual operation, the following happens:
●
All front panel keys are enabled.
●
The main softkey menu of the current mode is displayed.
●
The measurement diagrams, traces and display fields are displayed again.
●
If, at the time of pressing the "LOCAL" softkey, the synchronization mechanism via
*OPC, *OPC? or *WAI is active, the currently running measurement procedure is
Manuel d'utilisation 1179.5860.09 - 12
Device-specific error code
Brief description of the error
Time the message occurred
on page 978
on page 977
Remote Display Update
Network operation and remote control
Network and remote control settings
softkey are displayed.
587
587