Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt <Frequency>
This command defines the start frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the sweep list covers, the R&S FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
greater than the span the sweep list covers, the R&S FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
Suffix:
<ri>
Parameters:
<Frequency>
Example:
Manual operation:
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP <Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the sweep list covers, the R&S FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
greater than the span the sweep list covers, the R&S FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
Suffix:
<ri>
Parameters:
<Frequency>
Example:
Manual operation:
Manuel d'utilisation 1179.5860.09 - 12
Configuring and performing measurements
LIST:RANG2:DET AVER
See
"Detector"
on page 211
1..n
Selects the measurement range.
Numeric value.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
LIST:RANG2:STAR 2MHZ
See
"Range Start / Range Stop"
1..n
Selects the measurement range.
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
LIST:RANG2:STOP 5MHZ
See
"Range Start / Range Stop"
Remote commands
on page 211
on page 211
690
690