<li>
Return values:
<LowerACH>
<UpperACH>
Example:
Usage:
Manual operation:
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe <State>[, <State>]
This command turns the relative limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
<li>
Parameters:
<State>
<State>
Manual operation:
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute <LowerLimit>[,
This command defines the absolute limit of the alternate channels.
Manuel d'utilisation 1179.5860.09 - 12
irrelevant
text value
Thestate of the limit check for the lower adjacent channels.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
text value
The state of the limit check for the upper adjacent channels.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Query only
See
"Limit Check"
irrelevant
irrelevant
ON | OFF | 1 | 0
Relative limit check for lower adjacent channel
*RST:
ON | OFF | 1 | 0
Relative limit check for upper adjacent channel
*RST:
See
"Limit Check"
<UpperLimit>]
Configuring and performing measurements
on page 146
0
0
on page 146
Remote commands
647